EU Authorized Representative for Ventilator and CPAP Manufacturers: Rules 9 and 12, Class IIa and IIb

Lung ventilators are class IIb under Rules 9 and 12 of Annex VIII to Regulation (EU) 2017/745, and MDCG 2021-24 states that all lung ventilators should be in that class. A CPAP device for sleep apnoea without a monitoring function is class IIa. Either way a notified body is involved, and a manufacturer established outside the EU must designate a sole authorized representative under Article 11(1) before the first unit is placed on the Union market.

Which products this page covers

This page covers active respiratory support equipment: intensive care and transport lung ventilators, home mechanical ventilators, anaesthesia ventilators, continuous positive airway pressure (CPAP) and bilevel devices for sleep apnoea, and the breathing circuits, masks and humidifiers used with them. All of these are active devices under Article 2(4) of the MDR, and a ventilator, which administers kinetic energy to the lungs, is an active therapeutic device under Section 2.4 of Annex VIII.

Rule 9 and Rule 12: how ventilators are classified

Two rules of Annex VIII reach the same result for a lung ventilator. Rule 9 looks at the energy administered to the body:

"All active therapeutic devices intended to administer or exchange energy are classified as class IIa unless their characteristics are such that they may administer energy to or exchange energy with the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are classified as class IIb."

Source: Regulation (EU) 2017/745, Annex VIII, Rule 9

Rule 12 looks at the substances, here breathing gas, administered to the body:

"All active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are classified as class IIa, unless this is done in a manner that is potentially hazardous, taking account of the nature of the substances involved, of the part of the body concerned and of the mode of application in which case they are classified as class IIb."

Source: Regulation (EU) 2017/745, Annex VIII, Rule 12

MDCG 2021-24 resolves the question of whether a ventilator is potentially hazardous in its note on Rule 9:

"For instance, all devices intended to emit ionizing radiation, all lung ventilators and lithotriptors should be in Class IIb."

Source: MDCG 2021-24, Guidance on classification of medical devices, Rule 9, Note 1

The guidance lists lung ventilators under the class IIb branch of Rule 9 (kinetic energy) and ventilators and anaesthesia machines under the class IIb branch of Rule 12. The same guidance places sleep apnoea ventilators without monitoring function under the class IIa branch of Rule 9. This is the fact that decides the class of a CPAP device: a CPAP unit that only delivers positive airway pressure is class IIa, while a device that also monitors vital parameters, or a ventilator that supports or replaces breathing, is class IIb.

Where a device has a monitoring function alongside its therapeutic function, Section 3.5 of Annex VIII applies the stricter rule.

Accessories: circuits, masks and humidifiers

Accessories are classified in their own right under Section 3.2 of Annex VIII. Breathing circuits and tubing fall under Rule 2 as non-invasive devices channelling gases for introduction into the body:

"if they may be connected to a class IIa, class IIb or class III active device; or"

Source: Regulation (EU) 2017/745, Annex VIII, Rule 2, first indent

Tracheostomy or tracheal tubes connected to a ventilator are invasive with respect to a body orifice and connected to a class IIb active device; the last sentence of Rule 5 places such devices in class IIa. MDCG 2021-24 gives tracheostomy or tracheal tubes connected to a ventilator as a class IIa example under that sentence. A CPAP mask that rests on intact skin and is not itself invasive falls to Rule 1 or Rule 2 depending on whether the manufacturer intends it to channel gas to the patient; the intended purpose written in the instructions for use decides.

Classification table

Product variantClassRuleNotified body
Lung ventilator (intensive care, transport, home)IIbAnnex VIII, Rules 9 and 12Yes, Annex IX or Annex X plus XI
CPAP device for sleep apnoea without monitoring functionIIaAnnex VIII, Rule 9, first branchYes
CPAP or bilevel device with monitoring of vital parametersIIbAnnex VIII, Rules 9, 10 and 12; Section 3.5Yes
Breathing circuit or tubing connected to a ventilatorIIaAnnex VIII, Rule 2, first indentYes
Tracheal tube connected to a ventilatorIIaAnnex VIII, Rule 5, last sentenceYes

Notified body involvement

Every row in the table is class IIa or above, so a notified body is involved in each case. For class IIb:

"Manufacturers of class IIb devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of Annex IX, and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device per generic device group."

Source: Regulation (EU) 2017/745, Article 52(4)

For a class IIa CPAP device the route is Article 52(6): assessment of the quality management system under Chapters I and III of Annex IX with an assessment of the technical documentation of at least one representative device for each category of devices.

The authorized representative: Article 11

Whether a ventilator or CPAP manufacturer needs an EU authorized representative does not depend on the class of the product. It depends on where the manufacturer is established. Article 11(1) is the operative sentence:

"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative."

Source: Regulation (EU) 2017/745, Article 11(1)

Article 11(2) requires the mandate to be accepted in writing and to cover at least all devices of the same generic device group. Article 11(3) then lists the minimum tasks. For a ventilator manufacturer the first task, verifying that the notified body procedure has been carried out, means that Medex checks the certificate before accepting the mandate.

"(a) verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer;"

Source: Regulation (EU) 2017/745, Article 11(3)(a)

"(b) keep available a copy of the technical documentation, the EU declaration of conformity and, if applicable, a copy of the relevant certificate, including any amendments and supplements, issued in accordance with Article 56, at the disposal of competent authorities for the period referred to in Article 10(8);"

Source: Regulation (EU) 2017/745, Article 11(3)(b)

"(g) immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated;"

Source: Regulation (EU) 2017/745, Article 11(3)(g)

One mandate for the European Union and Türkiye

Medex is established in Türkiye and registered in EUDAMED as an authorized representative under SRN TR-AR-000057550, so a ventilator or CPAP manufacturer outside both the EU and Türkiye covers both markets with one mandate. MDCG 2022-16, footnote 1:

"The ’Union market’ refers to the territories of the European Union Member States, and due to the European Economic Area (EEA) is extended to Norway, Lichtenstein and Iceland, and via the Customs Union Agreement to Turkey."

Source: MDCG 2022-16, Guidance on Authorised Representatives, footnote 1

The Commission's Notice to Stakeholders of March 2022 draws the conclusion:

"manufacturers established outside the territory of the EU or Turkey only need to designate one authorised representative, either in the EU or in Turkey, in order to place medical devices on the market in the EU or in Turkey."

Source: European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices, Section I.2

Placing the product on the Turkish market also requires registration in ÜTS (Ürün Takip Sistemi), Türkiye's product tracking system, a separate service under a separate contract: see ÜTS registration for foreign manufacturers and ventilator ÜTS registration in Türkiye (Turkish).

What Medex does under the mandate

We verify that the EU declaration of conformity and technical documentation for each ventilator or CPAP device exist and that the notified body procedure has been carried out, keep copies for the period in Article 10(8), register in EUDAMED and check your Basic UDI-DI records, answer competent authority requests, cooperate on corrective actions and pass complaints and suspected incidents to you without delay. A person responsible for regulatory compliance is available in-house.

A class IIb ventilator falls in the band of €1,200 to 1,600 per year and a class IIa CPAP device in the band of €1,000 to 1,300 per year, fixed for three years from signature. Setup is €0 and a change of representative costs nothing; see pricing and EU authorized representative cost. We do not write the technical documentation or the clinical evaluation for respiratory equipment. We are not a notified body and we do not find importers or distributors. Offices in Ankara and İstanbul.

The symbol that identifies the representative on your label is explained in our guide to the EC REP meaning. If some devices are still placed on the market under the former directives, see the authorized representative for legacy devices.

Related pages

See also defibrillators, ultrasound and bone densitometers and blood pressure monitors and thermometers; all groups are on the medical device categories page. The general guide is EU authorized representative; the MDR classification tool walks through Annex VIII.

Frequently asked questions

Is a CPAP device class IIa or class IIb?

MDCG 2021-24 lists sleep apnoea ventilators without monitoring function as a class IIa example under Rule 9. A device that also monitors vital physiological parameters, or that supports or replaces breathing as a lung ventilator does, is class IIb. The intended purpose in the instructions for use decides.

Why are lung ventilators class IIb and not class IIa?

Rule 9 places active therapeutic devices in class IIb when they administer energy in a potentially hazardous way. MDCG 2021-24 states in its note on Rule 9 that all lung ventilators should be in class IIb. Rule 12 reaches the same result for the breathing gas administered.

Do we need a separate authorized representative for breathing circuits and masks?

No. One mandate with Medex can list the ventilator and its accessories; each accessory is classified in its own right under Section 3.2 of Annex VIII.

Does Medex check our notified body certificate?

Yes. Article 11(3)(a) requires the authorized representative to verify that the conformity assessment procedure has been carried out, and Article 11(3)(b) requires a copy of the certificate to be kept for competent authorities.

Does one mandate cover the EU and Türkiye for a ventilator manufacturer?

Yes. The Commission Notice of March 2022 states that manufacturers established outside the EU or Turkey only need to designate one authorized representative, either in the EU or in Turkey. Medex is established in Türkiye and registered in EUDAMED as an authorized representative.

Sources and further reading

Regulation (EU) 2017/745 on medical devices, consolidated text: eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20250110

MDCG 2021-24, Guidance on classification of medical devices: health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en

MDCG 2022-16, Guidance on Authorised Representatives Regulation (EU) 2017/745 and Regulation (EU) 2017/746: health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en

European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices (March 2022): health.ec.europa.eu/system/files/2022-04/md_eu-turkey_customs-union_en.pdf

Request a fixed-price EU authorized representative quote for your ventilator or CPAP device.

Request a proposal  ·  View pricing  ·  Appoint Medex as your authorized representative