Label, IFU, symbols and languages checked against Annex I, Section 23
We check your label and your instructions for use against Section 23 of Annex I, the symbols you use and the language rules of each market you sell in. You get the label back marked up, with every change tied to the requirement behind it. To start, send us the label artwork, the instructions for use and the list of markets.
Who this is for
Manufacturers outside the EU whose label has to name their authorised representative.
Manufacturers of self-declared Class I and Class A devices, where no notified body reads the label before the device reaches the market.
Manufacturers changing a label, adding a market or adding a language.
Manufacturers registering devices in Türkiye, where the label and the instructions for use are part of the registration file.
What the Regulation requires
The language rule sits in the manufacturer's general obligations:
"Manufacturers shall ensure that the device is accompanied by the information set out in Section 23 of Annex I in an official Union language(s) determined by the Member State in which the device is made available to the user or patient. The particulars on the label shall be indelible, easily legible and clearly comprehensible to the intended user or patient." (MDR Article 10(11))
Section 23.2 lists what the label carries, including a line that labels sometimes leave out:
"an indication that the device is a medical device. If the device is intended for clinical investigation only, the words 'exclusively for clinical investigation';" (MDR Annex I, 23.2(q))
Instructions for use come with every device, with one exception:
"Instructions for use shall be provided together with devices. By way of exception, instructions for use shall not be required for class I and class IIa devices if such devices can be used safely without any such instructions and unless otherwise provided for elsewhere in this Section." (MDR Annex I, 23.1(d))
Where an importer brings the device into the Union, the importer's details have to appear as well:
"Importers shall indicate on the device or on its packaging or in a document accompanying the device their name, registered trade name or registered trade mark, their registered place of business and the address at which they can be contacted, so that their location can be established." (MDR Article 13(3))
What we check on the label
We go through Section 23.2 point by point and record, for each point, whether it applies, where it appears on the label and whether it is correct.
Device name, the details needed to identify the device and, where it is not obvious, the intended purpose.
The manufacturer's name and address and, for a manufacturer outside the Union, the authorised representative's name and address.
Lot or serial number, the use-by date or the date of manufacture.
The UDI carrier, in machine-readable and human-readable form.
Storage and handling conditions, warnings and precautions.
Single-use, sterile state and sterilisation method, where they apply.
The indication that the device is a medical device.
The CE marking, and the notified body number where a notified body is involved.
Our guide to medical device labelling requirements explains each point in more detail.
Symbols
Much of a label is carried by symbols, and each symbol has to mean what the label needs it to mean. We check every symbol against ISO 15223-1, the standard for symbols used with information supplied by the manufacturer: the MD symbol, the authorised representative symbol, the manufacturer, the date of manufacture, the UDI and the importer symbol among them. Our guide to medical device symbols lists them. For the change from the EC REP symbol to the amended symbol, our free EU REP label transition checklist (PDF) lists the label, artwork and documentation points to update.
What we check in the instructions for use
Section 23.4 sets the content of the instructions for use. We check that the intended purpose, indications, contraindications, target patient group and intended users are stated, that residual risks and warnings match the risk file, and that the particulars shared with the label read the same in both. Where the instructions for use and the label disagree on the device name, the variants or the intended purpose, we mark both.
Languages
The label and the instructions for use have to be in the language or languages set by each Member State where the device is made available. We build a language table from your list of markets and check that each market's language is covered. Our guide on label language requirements explains the rules by market.
Labels for the Turkish market
From our registration work in ÜTS, the Turkish national database, labels and product images have been sent back for reasons that are easy to avoid:
The MD symbol was missing from the label.
The CE marking was missing from the instructions for use.
The product image showed a UDI that belonged to another product.
The label carried the details of another model.
We check for each of these before the file goes to ÜTS.
Process
You send the label artwork, the instructions for use, the device list and the list of markets.
We review the label against Section 23.2, the instructions for use against Section 23.4, the symbols and the languages.
We send the marked-up label and the review table.
Your designer updates the artwork. We check the updated version.
What you receive
The label marked up, with each change tied to the point of Section 23 behind it.
A review table for the instructions for use.
The list of symbols used, with each symbol's name.
The language table for your markets.
What we need from you
Label artwork for each packaging level, as editable or high-resolution files.
The instructions for use as they stand.
The device list with risk class and Basic UDI-DI.
The list of markets where the device is made available.
The EU declaration of conformity, so the names and codes can be compared.
Out of scope
Notified body conformity assessment and certification.
Testing and laboratory work, including choosing a laboratory or arranging tests.
Interpreting or changing your technical specifications.
Finding importers or distributors.
Related services
FAQs
Does the label have to show the authorised representative?
Yes, when the manufacturer's registered place of business is outside the Union. Annex I, Section 23.2(d) requires the authorised representative's name and the address of its registered place of business.
Do Class I devices need instructions for use?
Usually yes. Annex I, Section 23.1(d) allows an exception for Class I and Class IIa devices only where they can be used safely without instructions and nothing else in Section 23 requires them.
Do you redesign the label artwork?
No. We mark up the label and tell you what has to change and why. Your designer makes the change and we check the new version.
Which symbol standard do you check against?
ISO 15223-1, the standard for symbols used with information supplied by the manufacturer.
Do you check the importer's details?
Yes. Article 13(3) of the MDR requires the importer's name and address on the device, its packaging or an accompanying document, and the importer's label must not cover the manufacturer's information.
Do you review IVD labels as well?
Yes. We review them against the IVDR's own Annex I.