EU Authorized Representative for Blood Pressure Monitor and Thermometer Manufacturers: Rule 1 with Measuring Function or Rule 10
A mechanical sphygmomanometer or a non-electronic thermometer is a class I device with a measuring function under Rule 1 of Annex VIII to Regulation (EU) 2017/745, and Article 52(7)(b) limits the notified body to the metrological requirements. An electronic blood pressure monitor or thermometer is an active device and class IIa under Rule 10. In both cases a manufacturer outside the EU must designate a sole authorized representative under Article 11(1).
Which products this page covers
This page covers devices that measure arterial blood pressure and body temperature. The group splits along one line: whether the device is active. An electronic upper-arm or wrist blood pressure monitor and an electronic or infrared thermometer are active devices. A mechanical aneroid sphygmomanometer and a liquid-in-glass or other non-electronic clinical thermometer are not, because their operation does not depend on a source of energy other than that generated by the human body or by gravity (Article 2(4)).
Non-active instruments: Rule 1 and the measuring function
A non-active sphygmomanometer or thermometer touches only intact skin and is covered by the fall-back rule:
"All non-invasive devices are classified as class I, unless one of the rules set out hereinafter applies."
Source: Regulation (EU) 2017/745, Annex VIII, Rule 1
Class I is not the end of the story, because these instruments have a measuring function. Article 52(7) lets a class I manufacturer issue its own EU declaration of conformity after drawing up the technical documentation of Annexes II and III, but for devices with a measuring function it requires the procedures of Chapters I and III of Annex IX or Part A of Annex XI, with the notified body involvement limited:
"(b) in the case of devices with a measuring function, to the aspects relating to the conformity of the devices with the metrological requirements;"
Source: Regulation (EU) 2017/745, Article 52(7)(b)
MDCG 2021-24 sets out three criteria that, fulfilled together, indicate a measuring function: the device is intended to measure a physiological parameter quantitatively, the result is displayed in legal units or compared to a reference point in legal units, and the intended purpose implies accuracy where a failure of that accuracy could significantly harm the patient. A blood pressure reading in mmHg and a temperature in degrees Celsius meet all three. The metrological requirements the notified body checks are in Annex I, Section 15, which requires sufficient accuracy, precision and stability, stated limits of accuracy, and results expressed in legal units.
The result is a class I device with a measuring function (Im): the manufacturer issues the EU declaration of conformity itself, but a notified body certificate limited to the metrological aspects must exist before the CE mark is applied. See also class I medical devices.
Electronic monitors and thermometers: Rule 10
An electronic device that measures blood pressure or temperature is an active device intended for diagnosis and monitoring in the sense of Section 2.5 of Annex VIII, so Rule 10 applies:
"if they are intended to allow direct diagnosis or monitoring of vital physiological processes, unless they are specifically intended for monitoring of vital physiological parameters and the nature of variations of those parameters is such that it could result in immediate danger to the patient, for instance variations in cardiac performance, respiration, activity of the central nervous system, or they are intended for diagnosis in clinical situations where the patient is in immediate danger, in which cases they are classified as class IIb."
Source: Regulation (EU) 2017/745, Annex VIII, Rule 10, third indent
MDCG 2021-24 lists electronic thermometers and electronic blood pressure measuring equipment as class IIa examples under that indent. The class IIb branch is reserved for continuous surveillance:
"Vital physiological processes and parameters include, for example, respiration, heart rate, cerebral functions, blood gases, blood pressure and body temperature. Medical devices intended to be used for continuous surveillance of vital physiological parameters in anesthesia, intensive care or emergency care are in class IIb, whilst medical devices intended to be used to obtain readings of vital physiological signals as part of routine checkups or self-monitoring are in class IIa."
Source: MDCG 2021-24, Guidance on classification of medical devices, Rule 10, Note 3
A home blood pressure monitor or a forehead thermometer used for routine readings is therefore class IIa. A multi-parameter patient monitor that alarms on blood pressure or temperature in intensive care is class IIb.
Classification table
| Product variant | Class | Rule | Notified body |
|---|---|---|---|
| Aneroid (mechanical) sphygmomanometer | I with measuring function (Im) | Annex VIII, Rule 1; Article 52(7)(b) | Limited to metrological aspects |
| Non-electronic clinical thermometer | I with measuring function (Im) | Annex VIII, Rule 1; Article 52(7)(b) | Limited to metrological aspects |
| Electronic upper-arm or wrist blood pressure monitor | IIa | Annex VIII, Rule 10, third indent | Yes |
| Electronic or infrared thermometer | IIa | Annex VIII, Rule 10, third indent | Yes |
| Multi-parameter patient monitor with alarms for intensive care | IIb | Annex VIII, Rule 10, third indent, class IIb branch | Yes |
Notified body involvement
For the class IIa branch the notified body assesses the quality management system and the technical documentation of at least one representative device per category under Article 52(6). For the class I measuring branch its role is limited to the metrological requirements; the manufacturer still issues its own declaration of conformity.
The authorized representative: Article 11
Whether a blood pressure monitor or thermometer manufacturer needs an EU authorized representative does not depend on the class of the product. It depends on where the manufacturer is established. Article 11(1) is the operative sentence:
"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative."
Source: Regulation (EU) 2017/745, Article 11(1)
Article 11(2) requires the mandate to be accepted in writing and to cover at least all devices of the same generic device group. Article 11(3) then lists the minimum tasks. For class I (Im) devices the representative verifies the declaration of conformity and the metrological certificate; for class IIa devices it verifies the Annex IX certificate.
"(a) verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer;"
Source: Regulation (EU) 2017/745, Article 11(3)(a)
"(b) keep available a copy of the technical documentation, the EU declaration of conformity and, if applicable, a copy of the relevant certificate, including any amendments and supplements, issued in accordance with Article 56, at the disposal of competent authorities for the period referred to in Article 10(8);"
Source: Regulation (EU) 2017/745, Article 11(3)(b)
"(g) immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated;"
Source: Regulation (EU) 2017/745, Article 11(3)(g)
One mandate for the European Union and Türkiye
Medex is established in Türkiye and registered in EUDAMED as an authorized representative under SRN TR-AR-000057550, so a blood pressure monitor or thermometer manufacturer outside both the EU and Türkiye covers both markets with one mandate. MDCG 2022-16, footnote 1:
"The ’Union market’ refers to the territories of the European Union Member States, and due to the European Economic Area (EEA) is extended to Norway, Lichtenstein and Iceland, and via the Customs Union Agreement to Turkey."
Source: MDCG 2022-16, Guidance on Authorised Representatives, footnote 1
The Commission's Notice to Stakeholders of March 2022 draws the conclusion:
"manufacturers established outside the territory of the EU or Turkey only need to designate one authorised representative, either in the EU or in Turkey, in order to place medical devices on the market in the EU or in Turkey."
Source: European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices, Section I.2
Placing the product on the Turkish market also requires registration in ÜTS (Ürün Takip Sistemi), Türkiye's product tracking system, a separate service under a separate contract: see ÜTS registration for foreign manufacturers and blood pressure monitor ÜTS registration in Türkiye (Turkish).
A separate Turkish page exists for thermometer ÜTS registration in Türkiye (Turkish).
What Medex does under the mandate
We verify that the EU declaration of conformity and technical documentation for each blood pressure monitor or thermometer device exist and that the notified body procedure has been carried out, keep copies for the period in Article 10(8), register in EUDAMED and check your Basic UDI-DI records, answer competent authority requests, cooperate on corrective actions and pass complaints and suspected incidents to you without delay. A person responsible for regulatory compliance is available in-house.
A class I (Im) instrument falls in the band of €800 to 1,000 per year and a class IIa electronic monitor in the band of €1,000 to 1,300 per year, fixed for three years from signature. Setup is €0 and a change of representative costs nothing; see pricing and EU authorized representative cost. We do not perform the metrological testing and we do not act as the notified body for the measuring function. We are not a notified body and we do not find importers or distributors. Offices in Ankara and İstanbul.
The symbol that identifies the representative on your label is explained in our guide to the EC REP meaning. If some devices are still placed on the market under the former directives, see the authorized representative for legacy devices.
Related pages
See also ECG devices, continuous glucose monitors and ultrasound and bone densitometers; all groups are on the medical device categories page. The general guide is EU authorized representative.
Frequently asked questions
Is an electronic blood pressure monitor class I or class IIa?
Class IIa. It is an active device intended for diagnosis and monitoring, and Rule 10 places devices that allow direct diagnosis or monitoring of vital physiological processes in class IIa. MDCG 2021-24 lists electronic blood pressure measuring equipment as a class IIa example.
Can a thermometer be class I?
Yes, if it is not active. A non-electronic clinical thermometer is a non-invasive device under Rule 1, class I, with a measuring function. Article 52(7)(b) then limits the notified body to the metrological requirements. An electronic or infrared thermometer is active and class IIa under Rule 10.
What does a class I measuring function mean for the notified body?
The manufacturer still issues its own EU declaration of conformity, but Article 52(7) requires the procedures of Annex IX Chapters I and III or Annex XI Part A, with the notified body limited to the metrological requirements.
Our home monitor also stores readings in an app. Does that change the class?
The app is classified in its own right under Section 3.2 of Annex VIII. Software that analyses readings for diagnostic decisions falls under Rule 11; see our medical device software page.
Does one Medex mandate cover both the EU and Türkiye?
Yes. MDCG 2022-16 notes that the Union market extends via the Customs Union Agreement to Turkey, and the Commission Notice of March 2022 confirms that one authorized representative, in the EU or in Türkiye, is enough.
Sources and further reading
Regulation (EU) 2017/745 on medical devices, consolidated text: eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20250110
MDCG 2021-24, Guidance on classification of medical devices: health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
MDCG 2022-16, Guidance on Authorised Representatives Regulation (EU) 2017/745 and Regulation (EU) 2017/746: health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices (March 2022): health.ec.europa.eu/system/files/2022-04/md_eu-turkey_customs-union_en.pdf