EU Authorized Representative for Syringe and Needle Manufacturers: Rules 2 and 6, Measuring Function and the Mandate
Syringes and needles sit on both sides of the class I line in Regulation (EU) 2017/745. An empty syringe without a needle is a non-invasive class I device under Rule 2 of Annex VIII, while a syringe with a needle or a hypodermic needle is a surgically invasive device for transient use and class IIa under Rule 6, with spinal needles in class III. A manufacturer established outside the Union must designate a sole authorized representative under Article 11 for all of them, and one mandate covers the EU and Türkiye.
A syringe is not an active device
"For instance, energy generated by human muscle and applied to the plunger of a syringe (thus causing a substance to be delivered to a patient) does not make this syringe an active device." (MDCG 2021-24, Section 3.1, active devices)
A manual syringe is therefore classified under the non-invasive and invasive rules, not under Rules 9 to 13. Only a powered injector or syringe driver is active, and those fall under Rule 12 (see the infusion sets page).
Empty syringes without needles: Rule 2
A syringe supplied without a needle does not penetrate the body. It channels a liquid for eventual administration, so Rule 2 applies:
"All non-invasive devices intended for channelling or storing blood, body liquids, cells or tissues, liquids or gases for the purpose of eventual infusion, administration or introduction into the body are classified as class IIa:" (MDR Annex VIII, Rule 2)
- "if they may be connected to a class IIa, class IIb or class III active device; or"
- "In all other cases, such devices are classified as class I."
MDCG 2021-24 lists empty syringes without needles among the class I examples under the last sentence, and syringes for infusion pumps among the class IIa examples under the first indent, because an infusion pump is a class IIb active device. The intended purpose on the label decides which line applies. A syringe sold sterile stays class I, with the notified body involved for the sterile aspects only.
Syringes with needles and hypodermic needles: Rule 6
Once a needle is attached, the product penetrates the skin outside a body orifice. Section 2.2(b) of Annex VIII defines as surgically invasive "a device which produces penetration other than through a body orifice.", and an injection takes seconds, so Rule 6 applies:
"All surgically invasive devices intended for transient use are classified as class IIa unless they:" (MDR Annex VIII, Rule 6)
- "are intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III;"
- "are intended to administer medicinal products by means of a delivery system, if such administration of a medicinal product is done in a manner that is potentially hazardous taking account of the mode of application, in which case they are classified as class IIb."
The class IIa examples in MDCG 2021-24 include needles used for suturing, needles or syringes, and lancets. Two exceptions move specific products upward. Spinal needles are listed as class III under the central nervous system indent, because Section 2.7 defines that system as the brain, meninges and spinal cord. Refillable insulin pens are listed as class IIb under the medicinal product delivery indent. A blood collection needle is a transient surgically invasive device and stays in class IIa unless it is intended for use in the central circulatory system.
Graduated syringes and the measuring function
"Spoons or plastic syringes co-packed with medicinal products and used to measure a quantity of that medicinal product to be administered to the patient are in this category." (MDCG 2021-24, Section 3.1, devices with a measuring function)
For a class I syringe with a measuring function Article 52(7) brings in the notified body, but limits its involvement:
"in the case of devices with a measuring function, to the aspects relating to the conformity of the devices with the metrological requirements;" (MDR Article 52(7)(b))
A class IIa syringe with needle is assessed under Article 52(6) anyway, so the measuring function does not change its route. An oral dosing syringe co-packed with a medicine is listed by MDCG 2021-24 under Rule 5 as class I because it is used in the oral cavity.
Prefilled syringes
"However, if the device intended to administer a medicinal product and the medicinal product are placed on the market in such a way that they form a single integral product which is intended exclusively for use in the given combination and which is not reusable, that single integral product shall be governed by Directive 2001/83/EC or Regulation (EC) No 726/2004, as applicable." (MDR Article 1(9), second subparagraph)
The pharmaceutical company holds the marketing authorisation and the Annex I requirements apply to the syringe part. An empty syringe sold to that company is a device in its own right and needs CE marking, a declaration of conformity and, for a manufacturer outside the Union, an authorized representative.
Classification table for syringes and needles
| Product variant | Class | Rule | Notified body |
|---|---|---|---|
| Empty syringe without needle, sterile, manual use | I (sterile) | Rule 2, last sentence | Limited to sterility, Article 52(7)(a) |
| Syringe for infusion pumps or syringe drivers | IIa | Rule 2, first indent | Yes |
| Syringe with needle, hypodermic needle, lancet, blood collection needle | IIa | Rule 6 | Yes |
| Spinal needle | III | Rule 6, central nervous system indent | Yes |
| Refillable insulin pen | IIb | Rule 6, medicinal product delivery indent | Yes |
| Oral dosing syringe co-packed with a medicine | I, measuring function | Rule 5, oral cavity exception | Limited to metrology, Article 52(7)(b) |
| Prefilled syringe forming a single integral product with the medicine | Not a device | Article 1(9) | Medicinal product rules |
Conformity assessment for a class IIa syringe and needle range
"Manufacturers of class IIa devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of Annex IX, and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device for each category of devices." (MDR Article 52(6))
The notified body samples at least one representative device per category, so a range with hypodermic syringes, spinal needles and insulin pens is sampled in each class. Sterility, biocompatibility, needle penetration and any sharps protection feature are documented in the technical documentation.
What the authorized representative mandate covers for syringes and needles
The appointment turns on where the manufacturer is established, not on the class of the device:
"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative." (MDR Article 11(1))
Article 11(3) fixes the minimum tasks:
"The mandate shall require, and the manufacturer shall enable, the authorised representative to perform at least the following tasks in relation to the devices that it covers:" (MDR Article 11(3))
- "verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer;"
- "comply with the registration obligations laid down in Article 31 and verify that the manufacturer has complied with the registration obligations laid down in Articles 27 and 29;"
Syringes and needles ship under many catalogue numbers. The fee on our pricing page follows the number of Basic UDI-DIs, not catalogue numbers, so several hundred references under a few Basic UDI-DIs do not multiply the cost.
One mandate for the European Union and Türkiye
Medex Kurumsal Danışmanlık acts as EU authorized representative under EUDAMED SRN TR-AR-000057550, with offices in Ankara and İstanbul. A manufacturer of syringes and needles established outside the EU and Türkiye does not need two representatives:
"manufacturers established outside the territory of the EU or Turkey only need to designate one authorised representative, either in the EU or in Turkey, in order to place medical devices on the market in the EU or in Turkey." (European Commission, Notice to Stakeholders, EU-Turkey Customs Union Agreement in the field of medical devices, March 2022, section I.2)
"The ’Union market’ refers to the territories of the European Union Member States, and due to the European Economic Area (EEA) is extended to Norway, Lichtenstein and Iceland, and via the Customs Union Agreement to Turkey." (MDCG 2022-16, footnote 1)
Our mandate covers the EU Member States and Türkiye; we are not a Swiss representative or UK Responsible Person, not a notified body, and we do not find importers or distributors. The published bands on our pricing page are €800 to 1,000 per year for class I, €1,000 to 1,300 per year for class IIa, €1,500 to 2,000 per year for class III, with a €0 setup fee and rates fixed for three years from signature. Registration of the same syringes and needles in ÜTS (Ürün Takip Sistemi), Türkiye's product tracking system, is a separate service: ÜTS registration of syringes in Türkiye (Turkish).
The symbol that identifies the representative on your label is explained in our guide to the EC REP meaning. If some devices are still placed on the market under the former directives, see the authorized representative for legacy devices.
Related device categories and country pages
Related consumable groups: infusion sets, catheters and examination and surgical gloves. All groups are listed on the medical device categories page, and the manufacturer country pages start from the EU authorized representative by country list.
Frequently asked questions
Is a sterile empty syringe class I?
Yes. An empty syringe without a needle channels a liquid for administration and is class I under the last sentence of Rule 2, unless it is intended for use with an infusion pump, which makes it class IIa. Supplied sterile, the notified body is involved only for the sterile aspects under Article 52(7)(a).
Why is a syringe with a needle class IIa?
The needle produces penetration other than through a body orifice, which makes the product a surgically invasive device under Annex VIII, Section 2.2(b). An injection is transient use, so Rule 6 applies and the default class is IIa.
Which class is a spinal needle?
Class III. Rule 6 moves a transient surgically invasive device to class III where it is intended specifically for use in direct contact with the central nervous system, and MDCG 2021-24 lists spinal needles under that indent.
Does the scale on a syringe create a measuring function?
MDCG 2021-24 treats plastic syringes co-packed with medicinal products and used to measure a quantity of that product as devices with a measuring function. For a class I syringe this brings in the notified body for the metrological aspects only, under Article 52(7)(b).
Does one mandate cover the European Union and Türkiye?
Yes. The Commission Notice of March 2022 confirms that a manufacturer established outside the EU and Türkiye only needs to designate one authorized representative, either in the EU or in Türkiye. Medex is established in Türkiye under EUDAMED SRN TR-AR-000057550.
Sources and further reading
Regulation (EU) 2017/745 on medical devices (MDR), Articles 2, 10, 11, 18, 32, 52 and 54, Annex VIII - eur-lex.europa.eu
MDCG 2021-24, Guidance on classification of medical devices (October 2021) - health.ec.europa.eu
MDCG 2022-16, Guidance on authorised representatives (Regulation (EU) 2017/745 and Regulation (EU) 2017/746) - health.ec.europa.eu
European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices (March 2022) - health.ec.europa.eu