top of page
Medex EU Authorised Representative logo

EU Authorized Representative for Orthopedic Implant Manufacturers: Rule 8, Implant Card and the Mandate

Every orthopaedic implant is classified under Rule 8 of Annex VIII to Regulation (EU) 2017/745. Plates, screws, nails and bone cement are class IIb; total or partial joint replacements and implants in contact with the spinal column are class III, with screws, wedges and plates excepted. All of them carry the implant obligations of Articles 18, 32 and 10(8). A manufacturer established outside the Union must designate a sole authorized representative under Article 11, and one mandate covers the EU and Türkiye.

Rule 8 governs every orthopaedic implant

Plates, screws, nails, cages and prostheses remain after the procedure, so they are implantable devices under Article 2(5), even when a later operation removes them:

"Thus a plate used to fix a fractured bone and which remains in place after the procedure for at least 30 days is an implant even if it is taken out after the fracture has healed." (MDCG 2021-24, Section 3.1, implantable devices)

"All implantable devices and long-term surgically invasive devices are classified as class IIb unless they:" (MDR Annex VIII, Rule 8)

  • "have a biological effect or are wholly or mainly absorbed, in which case they are classified as class III;"
  • "are total or partial joint replacements, in which case they are classified as class III, with the exception of ancillary components such as screws, wedges, plates and instruments; or"
  • "are spinal disc replacement implants or are implantable devices that come into contact with the spinal column, in which case they are classified as class III with the exception of components such as screws, wedges, plates and instruments."

Trauma and fixation implants: class IIb

MDCG 2021-24 lists plates and pedicle screws as class IIb examples under the general rule, and plain bone cement stays there as well. Two points keep fixation hardware in class IIb even on the spine: the ninth indent excepts "components such as screws, wedges, plates and instruments" (MDR Annex VIII, Rule 8) from class III, and MDCG 2021-24 keeps bone cement out of the chemical change indent:

"Note 4: The clause about chemical change under this rule does not apply to products such as bone cements where the chemical change takes place during the placement and does not continue in long term." (MDCG 2021-24, Rule 8, Note 4)

Joint replacements and spinal implants: class III

The eighth indent moves total or partial joint replacements to class III; MDCG 2021-24 lists hip, knee, shoulder and ankle. The exception for ancillary components means the screws fixing an acetabular cup are class IIb while the cup is class III. The ninth indent does the same for the spine: MDCG 2021-24 lists spinal disc replacement implants, interbody fusion devices and hooks or stems in contact with the spinal column as class III. A pedicle screw is therefore class IIb; the cage it supports is class III.

Coatings, antibiotic cement and animal-derived grafts

Under the biological effect indent, MDCG 2021-24 lists biodegradable bone cements and implants claimed to be bioactive through surface coatings as class III. Antibiotic-loaded cement falls under Rule 14:

"All devices incorporating, as an integral part, a substance which, if used separately, can be considered to be a medicinal product, as defined in point 2 of Article 1 of Directive 2001/83/EC, including a medicinal product derived from human blood or human plasma, as defined in point 10 of Article 1 of that Directive, and that has an action ancillary to that of the devices, are classified as class III." (MDR Annex VIII, Rule 14)

MDCG 2021-24 lists bone cement with antibiotics as its first Rule 14 example; bone graft substitutes of animal origin are class III under Rule 18.

Implant card, SSCP and 15-year retention

"In addition, the manufacturer shall provide the information referred to in point (a) of the first subparagraph on an implant card delivered with the device." (MDR Article 18(1))

Article 18(3) removes screws, wedges, plates, wires, pins, clips and connectors from the card, but not nails, cages, prostheses or cement. The implant card must exist in the languages determined by each Member State where the implant is sold. Article 32 adds the summary of safety and clinical performance for every implantable device, and Article 10(8) sets the retention period that the authorized representative must match under Article 11(3)(b):

"Manufacturers shall keep the technical documentation, the EU declaration of conformity and, if applicable, a copy of any relevant certificate, including any amendments and supplements, issued in accordance with Article 56, available for the competent authorities for a period of at least 10 years after the last device covered by the EU declaration of conformity has been placed on the market. In the case of implantable devices, the period shall be at least 15 years after the last device has been placed on the market." (MDR Article 10(8))

Classification table for an orthopaedic portfolio

Product variantClassRuleNotified body
Bone plate, screw, intramedullary nail, K-wire, pinIIbRule 8, general ruleYes; screws, plates, wires and pins exempt from per-device TD assessment
Pedicle screw, rod and connector of a spinal fixation systemIIbRule 8, ninth indent exceptionYes
Interbody fusion cage, disc replacement, hook contacting the spinal columnIIIRule 8, ninth indentYes; Article 54 consultation
Hip, knee, shoulder or ankle prosthesisIIIRule 8, eighth indentYes; Article 54 consultation
Non-biodegradable bone cement without antibioticIIbRule 8, general rule; MDCG 2021-24 Note 4Yes
Bone cement with antibioticIIIRule 14Yes; medicines authority consultation
Bone graft substitute of animal originIIIRule 18Yes; Regulation (EU) No 722/2012

Conformity assessment: Article 52(3) and 52(4), Article 54 for class III implants

"Manufacturers of class III devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Annex IX. Alternatively, the manufacturer may choose to apply a conformity assessment as specified in Annex X coupled with a conformity assessment as specified in Annex XI." (MDR Article 52(3))

"However, for class IIb implantable devices, except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the assessment of the technical documentation as specified in Section 4 of Annex IX shall apply for every device." (MDR Article 52(4), second subparagraph)

A class IIb cage or nail is therefore assessed device by device, while screws and plates are sampled. For class III implants the notified body follows the Article 54 clinical evaluation consultation; see also post-market clinical follow-up.

What the authorized representative mandate covers for orthopaedic implants

The appointment turns on where the manufacturer is established, not on the class of the device:

"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative." (MDR Article 11(1))

Article 11(3) fixes the minimum tasks:

"The mandate shall require, and the manufacturer shall enable, the authorised representative to perform at least the following tasks in relation to the devices that it covers:" (MDR Article 11(3))

  • "verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer;"
  • "immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated;"

Article 11(5) makes the representative jointly and severally liable for defective devices where the manufacturer has not complied with Article 10, which is why implant mandates are reviewed closely before signature (see liability).

One mandate for the European Union and Türkiye

Medex Kurumsal Danışmanlık acts as EU authorized representative under EUDAMED SRN TR-AR-000057550, with offices in Ankara and İstanbul. A manufacturer of orthopaedic implants established outside the EU and Türkiye does not need two representatives:

"manufacturers established outside the territory of the EU or Turkey only need to designate one authorised representative, either in the EU or in Turkey, in order to place medical devices on the market in the EU or in Turkey." (European Commission, Notice to Stakeholders, EU-Turkey Customs Union Agreement in the field of medical devices, March 2022, section I.2)

"The ’Union market’ refers to the territories of the European Union Member States, and due to the European Economic Area (EEA) is extended to Norway, Lichtenstein and Iceland, and via the Customs Union Agreement to Turkey." (MDCG 2022-16, footnote 1)

Our mandate covers the EU Member States and Türkiye; we are not a Swiss representative or UK Responsible Person, not a notified body, and we do not find importers or distributors. The published bands on our pricing page are €1,200 to 1,600 per year for class IIb, €1,500 to 2,000 per year for class III, with a €0 setup fee and rates fixed for three years from signature. Registration of the same orthopaedic implants in ÜTS (Ürün Takip Sistemi), Türkiye's product tracking system, is a separate service: ÜTS registration of orthopaedic implants in Türkiye (Turkish).

The symbol that identifies the representative on your label is explained in our guide to the EC REP meaning. If some devices are still placed on the market under the former directives, see the authorized representative for legacy devices.

Related device categories and country pages

Related groups: reusable surgical instruments, surgical sutures and dental devices, which covers dental implants under the Rule 8 teeth indent. All groups are listed on the medical device categories page, and the manufacturer country pages start from the EU authorized representative by country list.

Frequently asked questions

Is a bone plate class IIb or class III?

Class IIb. Rule 8 places implantable devices in class IIb unless an indent applies, and MDCG 2021-24 lists plates as a class IIb example. Even a plate used on the spine stays in class IIb, because the ninth indent excepts screws, wedges, plates and instruments from class III.

Why is a knee prosthesis class III?

The eighth indent of Rule 8 classifies total or partial joint replacements in class III, with the exception of ancillary components such as screws, wedges, plates and instruments. MDCG 2021-24 lists hip, knee, shoulder and ankle as examples.

Do screws need an implant card?

No. Article 18(3) exempts screws, wedges, plates, wires, pins, clips and connectors. Nails, cages, prostheses and bone cement are not on that list and need the card.

Which class is bone cement?

Non-biodegradable cement without an active substance is class IIb under the general rule of Rule 8; MDCG 2021-24 Note 4 keeps it outside the chemical change indent. Biodegradable cement is class III under the biological effect indent, and cement with an antibiotic is class III under Rule 14.

Does one mandate cover the European Union and Türkiye?

Yes. Under the EU-Türkiye Customs Union a manufacturer established outside both territories needs only one authorized representative, established in the EU or in Türkiye. Medex is established in Türkiye under EUDAMED SRN TR-AR-000057550.

Sources and further reading

Regulation (EU) 2017/745 on medical devices (MDR), Articles 2, 10, 11, 18, 32, 52 and 54, Annex VIII - eur-lex.europa.eu

MDCG 2021-24, Guidance on classification of medical devices (October 2021) - health.ec.europa.eu

MDCG 2022-16, Guidance on authorised representatives (Regulation (EU) 2017/745 and Regulation (EU) 2017/746) - health.ec.europa.eu