EU Authorized Representative for Endoscope Manufacturers: Rules 5, 6 and 10 and the Mandate
Endoscopes introduced through a body orifice and connected to a light source are class IIa under the last sentence of Rule 5 of Annex VIII to Regulation (EU) 2017/745; laparoscopes are class IIa under Rule 6; neuro-endoscopes are class III. Processors and light sources are class IIa under Rule 10. A notified body is involved in every case, and a manufacturer outside the EU must designate a sole authorized representative under Article 11(1).
Which products this page covers
This page covers flexible and rigid endoscopes, video endoscopes and their processors and light sources, laparoscopes, arthroscopes, neuro-endoscopes, single-use endoscopes and endoscopic accessories such as biopsy forceps and fibre optics for surgical lasers. The group spans three rules of Annex VIII because an endoscope can enter the body through a natural orifice or through an incision, and because the optical and video parts are active while the insertion tube is not.
One definition sets the main branch. A surgically invasive device is:
"an invasive device which penetrates inside the body through the surface of the body, including through mucous membranes of body orifices with the aid or in the context of a surgical operation; and"
Source: Regulation (EU) 2017/745, Annex VIII, Section 2.2(a)
Rule 5: endoscopes introduced through a body orifice
A gastroscope, colonoscope, bronchoscope or cystoscope is invasive with respect to a body orifice and is connected to a light source or video processor, which is an active device. The last sentence of Rule 5 of Annex VIII governs that case:
"All invasive devices with respect to body orifices, other than surgically invasive devices, intended for connection to a class IIa, class IIb or class III active device, are classified as class IIa."
Source: Regulation (EU) 2017/745, Annex VIII, Rule 5
MDCG 2021-24 lists endoscopes using a light source in the visible spectrum and fibre optics in endoscopes connected to surgical lasers as class IIa examples under that sentence.
Rule 6: laparoscopes and other surgically invasive endoscopes
An endoscope introduced through an incision, such as a laparoscope, arthroscope or thoracoscope, is surgically invasive for transient use:
"All surgically invasive devices intended for transient use are classified as class IIa unless they:"
Source: Regulation (EU) 2017/745, Annex VIII, Rule 6
Class IIa is the default. The class III exception matters for neurosurgery:
"are intended specifically for use in direct contact with the heart or central circulatory system or the central nervous system, in which case they are classified as class III;"
Source: Regulation (EU) 2017/745, Annex VIII, Rule 6, third indent
MDCG 2021-24 lists neuro-endoscopes as a class III example under that indent. A rigid endoscope that is a reusable surgical instrument in the sense of Section 2.3 would be class I under the second indent of Rule 6 only if it has no connection to an active device; an endoscope connected to a light source or camera does not meet that definition.
Rule 10: light sources, processors and camera heads
The video processor, light source and camera head are active devices intended for diagnosis and monitoring:
"if they are intended to supply energy which will be absorbed by the human body, except for devices intended to illuminate the patient's body, in the visible spectrum, in which case they are classified as class I;"
Source: Regulation (EU) 2017/745, Annex VIII, Rule 10, first indent
Light delivered through the endoscope to image tissue is energy absorbed by the body for a diagnostic purpose, class IIa. The class I exception in that indent covers devices that illuminate the patient's body in the visible spectrum; MDCG 2021-24 explains in its note on illumination that a device used in the process of diagnosis is not limited to illumination and gives examination lamps and surgical microscopes as the class I examples. A video endoscopy system is intended for diagnosis and sits in class IIa.
Classification table
| Product variant | Class | Rule | Notified body |
|---|---|---|---|
| Flexible gastroscope, colonoscope or bronchoscope with light source | IIa | Annex VIII, Rule 5, last sentence | Yes |
| Single-use flexible endoscope through a body orifice | IIa | Annex VIII, Rule 5, last sentence | Yes |
| Rigid laparoscope, arthroscope, thoracoscope | IIa | Annex VIII, Rule 6 | Yes |
| Neuro-endoscope | III | Annex VIII, Rule 6, third indent | Yes |
| Video processor, light source, camera head | IIa | Annex VIII, Rule 10, first indent | Yes |
| Fibre optics in endoscopes connected to surgical lasers | IIa | Annex VIII, Rule 5, last sentence | Yes |
Notified body involvement
Every row is class IIa or above. The notified body route for class IIa:
"Manufacturers of class IIa devices, other than custom-made or investigational devices, shall be subject to a conformity assessment as specified in Chapters I and III of Annex IX, and including an assessment of the technical documentation as specified in Section 4 of that Annex of at least one representative device for each category of devices."
Source: Regulation (EU) 2017/745, Article 52(6)
For class III neuro-endoscopes Article 52(3) applies: conformity assessment under Annex IX, or type examination under Annex X with product conformity verification under Annex XI. Reusable endoscopes are reprocessed between patients, but the limited notified body involvement of Article 52(7)(c) applies only to class I reusable surgical instruments; a class IIa endoscope's reprocessing instructions are assessed within the ordinary Annex IX procedure.
The authorized representative: Article 11
Whether an endoscope manufacturer needs an EU authorized representative does not depend on the class of the product. It depends on where the manufacturer is established. Article 11(1) is the operative sentence:
"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative."
Source: Regulation (EU) 2017/745, Article 11(1)
Article 11(2) requires the mandate to be accepted in writing and to cover at least all devices of the same generic device group. Article 11(3) then lists the minimum tasks. For endoscopy systems the representative keeps one certificate that typically lists scopes, processors and accessories, and tracks every supplement as new models are added.
"(a) verify that the EU declaration of conformity and technical documentation have been drawn up and, where applicable, that an appropriate conformity assessment procedure has been carried out by the manufacturer;"
Source: Regulation (EU) 2017/745, Article 11(3)(a)
"(b) keep available a copy of the technical documentation, the EU declaration of conformity and, if applicable, a copy of the relevant certificate, including any amendments and supplements, issued in accordance with Article 56, at the disposal of competent authorities for the period referred to in Article 10(8);"
Source: Regulation (EU) 2017/745, Article 11(3)(b)
"(g) immediately inform the manufacturer about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated;"
Source: Regulation (EU) 2017/745, Article 11(3)(g)
One mandate for the European Union and Türkiye
Medex is established in Türkiye and registered in EUDAMED as an authorized representative under SRN TR-AR-000057550, so an endoscope manufacturer outside both the EU and Türkiye covers both markets with one mandate. MDCG 2022-16, footnote 1:
"The ’Union market’ refers to the territories of the European Union Member States, and due to the European Economic Area (EEA) is extended to Norway, Lichtenstein and Iceland, and via the Customs Union Agreement to Turkey."
Source: MDCG 2022-16, Guidance on Authorised Representatives, footnote 1
The Commission's Notice to Stakeholders of March 2022 draws the conclusion:
"manufacturers established outside the territory of the EU or Turkey only need to designate one authorised representative, either in the EU or in Turkey, in order to place medical devices on the market in the EU or in Turkey."
Source: European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices, Section I.2
Placing the product on the Turkish market also requires registration in ÜTS (Ürün Takip Sistemi), Türkiye's product tracking system, a separate service under a separate contract: see ÜTS registration for foreign manufacturers and endoscope ÜTS registration in Türkiye (Turkish).
What Medex does under the mandate
We verify that the EU declaration of conformity and technical documentation for each endoscope device exist and that the notified body procedure has been carried out, keep copies for the period in Article 10(8), register in EUDAMED and check your Basic UDI-DI records, answer competent authority requests, cooperate on corrective actions and pass complaints and suspected incidents to you without delay. A person responsible for regulatory compliance is available in-house.
A class IIa endoscope or processor falls in the band of €1,000 to 1,300 per year and a class III neuro-endoscope in the band of €1,500 to 2,000 per year, fixed for three years from signature. Setup is €0 and a change of representative costs nothing; see pricing and EU authorized representative cost. We do not validate reprocessing instructions or test optical performance. We are not a notified body and we do not find importers or distributors. Offices in Ankara and İstanbul.
The symbol that identifies the representative on your label is explained in our guide to the EC REP meaning. If some devices are still placed on the market under the former directives, see the authorized representative for legacy devices.
Related pages
Related surgical and imaging groups with their own pages include reusable surgical instruments, ultrasound and bone densitometers and aesthetic lasers and IPL. All groups are listed on the medical device categories page. The general guide is EU authorized representative.
Frequently asked questions
What class is a flexible endoscope under the MDR?
Class IIa. An endoscope introduced through a body orifice and connected to a light source or processor falls under the last sentence of Rule 5, and MDCG 2021-24 lists endoscopes using a light source in the visible spectrum as a class IIa example.
Is a laparoscope also class IIa?
Yes. It is surgically invasive for transient use and class IIa under Rule 6, unless it is intended for direct contact with the central nervous system, in which case the third indent of Rule 6 gives class III. MDCG 2021-24 lists neuro-endoscopes as class III.
Is the video processor classified separately from the scope?
Yes. Section 3.2 of Annex VIII requires devices used in combination to be classified separately. The processor and light source are active diagnostic devices supplying energy absorbed by the body and are class IIa under the first indent of Rule 10.
Can a rigid reusable endoscope be class I as a reusable surgical instrument?
Only if it has no connection to an active device, which is part of the definition in Section 2.3 of Annex VIII. An endoscope connected to a light source or camera does not meet that definition and is class IIa under Rule 5 or Rule 6.
Does one Medex mandate cover the EU and Türkiye for endoscopy systems?
Yes. The Commission Notice of March 2022 states that manufacturers outside the EU or Turkey only need to designate one authorized representative, either in the EU or in Turkey. Medex is established in Türkiye and registered in EUDAMED.
Sources and further reading
Regulation (EU) 2017/745 on medical devices, consolidated text: eur-lex.europa.eu/legal-content/EN/TXT/?uri=CELEX:02017R0745-20250110
MDCG 2021-24, Guidance on classification of medical devices: health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
MDCG 2022-16, Guidance on Authorised Representatives Regulation (EU) 2017/745 and Regulation (EU) 2017/746: health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en
European Commission, Notice to Stakeholders: EU-Turkey Customs Union Agreement in the field of medical devices (March 2022): health.ec.europa.eu/system/files/2022-04/md_eu-turkey_customs-union_en.pdf